4. Level correction
The conversion loss of the mixer can be taken into account both as a function of
frequency and with an average value. The frequency-dependent level correction
is used in this example for obtaining a higher precision. First select a table valid
for the selected band. As an alternative, level correction based on average value
can be used (see
“Level correction with average value” on page 4.286
5. Frequency-dependent level correction
➢
Activate a selection list which contains all tables stored on the hard disk of the
R&S FSP with
[
FREQUENCY
: EXTERNAL MIXER : SELECT BAND]
➢
Select the
CONV LOSS TABLE
field by means of the cursor keys.
➢
If a file with correction data already exists for the mixer used, select a table
valid with the cursor keys or the rotary knob and confirm with
ENTER
.
➢
If such a table is not yet available, create a new table with
[CONV LOSS TABLE : EDIT TABLE]
or copy from a disk to the hard disk of the R&S FSP with
[CONV LOSS TABLE : LOAD TABLE]
(See section
“Conversion Loss Tables” on page 4.271
and select the file created or loaded from a disk in the selection list as
described before.
A selected file is marked by
√
and will be drawn into the table SELECT BAND.
No further settings are necessary since the selected file contains all required
parameters.
➢
Quit the
SELECT BAND
menu using
[
ESC
] key
A span is automatically set which covers the whole V band (50 to 75 GHz).
➢
Set the frequency range to be examined using
[ FREQUENCY START : 52 GHz ]
and
[ FREQUENCY STOP : 60 GHz ]
➢
Reduce video bandwidth with
[
SWEEP COUPLING
: VIDEO BW MANUAL :
1 MHz
]
A correct signal identification using
AUTO ID
is thus possible (see also
“Remarks Concerning Signal Identification with AUTO ID” on page 4.278
The display shown in
is obtained.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...