4.9.5
Frequency-Converting Measurements
For frequency-converting measurements (e.g. on converters) the external generator
is able to set a constant frequency offset between the output frequency of the gener-
ator and the receive frequency of the R&S FSP and, in addition, the generator fre-
quency as a multiple of the R&S FSP.
Fig. 4.35 Test setup for frequency-converting measurements
FREQUENCY
OFFSET
The
FREQUENCY OFFSET
softkey activates the input of the frequency offset
between the output signal of the generator and the input frequency of the R&S FSP.
The value range depends on the selected generator.
The default setting is 0 Hz. Offsets <> 0 Hz are marked with the enhancement label
FRQ
.
If a positive frequency offset is entered, the tracking generator generates an output
signal above the receive frequency of the analyzer. In case of a negative frequency
offset it generates a signal below the receive frequency of the analyzer. The output
frequency of the generator is calculated as follows:
Generator frequency = receive fre frequency offset
Remote command:
SOUR:EXT:FREQ:OFFS 1GHZ
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...