R&S FSP
Instrument Functions
Basic Settings
4.148
Operating Manual 1164.4556.12 - 06
The
SELECTED LIMIT LINE
display field provides information concerning the char-
acteristics of the marked limit lines.
In the
LIMIT LINES
table, the limit lines compatible to the settings of the active
screen can be enabled.
New limit lines can be specified and edited in the
NEW LIMIT LINE and EDIT LIMIT
LINE
submenus, respectively.
The horizontal and vertical lines of the
DISPLAY LINES
submenu mark individual
levels or frequencies (span > 0) or times (span = 0) in the diagram.
The
SELECTED LIMIT LINE
table provides information about the characteristics of
the marked limit line:
The characteristics of the limit line are set in the
EDIT LIMIT LINE (=NEW LIMIT
LINE)
submenu.
SELECT LIMIT
LINE
The
SELECT LIMIT LINE
softkey activates the
LIMIT LINES
table and the selection
bar jumps to the uppermost name in the table.
The following information is offered in the columns of the table:
Name
name
Domain
frequency or time
Unit
vertical
scale
X-Axis
interpolation
Limit
upper/lower
limit
X-Scaling
absolute or relative frequencies/times
Y-Scaling
absolute or relative Y units
Threshold
absolute limit with relative Y units
Comment
commentary
Name
Enable the limit line.
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...