![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 172](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864172.webp)
R&S FSP
Instrument Functions
Analyzer Mode
4.100
Operating Manual 1164.4556.12 - 06
➢
Select the user standard you just created with
➢
Change the offset
➢
Save the user standard again to keep the current settings.
Remote command:
POW:ACH:PRES:RLEV:OFFS 10dB
SELECT TRACE
The
SELECT TRACE
softkey selects the trace on which the CP/ACP measurement
is to be performed. Only activated traces can be selected, i.e. traces not set to
BLANK.
Remote command:
SENS:POW:TRAC 1
Examples
Measurement of adjacent-channel power for a specific standard:
The adjacent-channel power is to be measured for a signal at 800 MHz with 0 dBm
level in line with IS136.
[
PRESET
]
Set the R&S FSP to the default setting.
[
FREQ
:
CENTER
: 800
MHz
]
Set the center frequency to 800 MHz.
[
AMPT
: 0
dBm
]
Set the reference level to 0 dBm.
[
MEAS
]
Call the menu for the measurement functions.
[
CHAN PWR / ACP
]
Select the channel and adjacent-channel power measure-
ment function. The measurement is performed with the
default settings or a previously defined setting. The sub-
menu for setting the desired new configuration is opened.
[
CP/ACP STANDARD
:
select IS136:
ENTER
]
Select the NADC (IS136) standard.
[
CP/ACP CONFIG
]
Call the submenu for configuration of the adjacent-channel
power measurement.
[
NO. OF ADJ CHAN
: 2
ENTER
]
Select two adjacent channels for the measurement, i.e.
the adjacent channel and the alternate channel are mea-
sured.
[
ADJUST SETTINGS
]
Set the optimum span, resolution bandwidth (RBW), video
bandwidth (VBW) and detector automatically for the mea-
surement. The absolute channel power and the relative
power of the adjacent channels are displayed on the
screen.
Change to the main menu for channel power measure-
ment.
[
ADJUST REF LVL
]
Set the reference level equal to the channel power mea-
sured.
PREV
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...