FREE RUN
The
FREE RUN
softkey activates the free-run sweep mode, i.e. start of a sweep is
not triggered. Once a measurement is completed, another is started immediately.
FREE RUN
is the default setting of R&S FSP.
Remote command:
TRIG:SOUR IMM
VIDEO
The
VIDEO
softkey activates triggering through the displayed voltage.
For the video triggering mode, a horizontal trigger line is shown in the diagram. It
may be used to set the trigger threshold between 0% and 100% of the overall dia-
gram height.
Remote command:
TRIG:SOUR VID
TRIG:LEV:VID 50 PCT
EXTERN
The
EXTERN
softkey activates triggering via a TTL signal at the input connector
EXT TRIGGER/GATE
on the rear panel.
The external trigger level can be adjusted in the range from 0.5 V to 3.5 V. The
default value is 1.4 V.
Remote command:
TRIG:SOUR EXT <numeric_value>
SWE:EGAT:SOUR EXT
TRIG:SEQ:LEV:EXT
IF POWER
The
IF POWER
softkey activates triggering of the measurement via signals which
are outside the measurement channel.
!
Side menu
TV TRIGGER ON | OFF
VERT SYNC
VERT SYNC ODD FIELD /
VERT SYNC EVEN FIELD
HOR SYNC
VIDEO POL POS | NEG
LINES 625 | 525
CCVS INT | EXT
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...