Remote Commands
R&S
®
Spectrum Rider FPH
319
User Manual 1321.1011.02 ─ 06.00
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<Mode>
FREQuency | CHANnel
FREQuency
Sets the marker frequency mode to frequency input (in Hz).
CHANnel
Sets the marker frequency mode to channel input (as a channel
number).
*RST:
FREQ
Example:
CALC:MARK:FREQ:MODE FREQ
Selects the frequency display mode.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:CENTer
This command matches the center frequency to the frequency of a marker.
If you use a delta marker, the R&S Spectrum Rider turns it into a normal marker.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK2:FUNC:CENT
Matches the center frequency to the frequency of marker 2.
Usage:
Event
CALCulate<1...2>:MARKer<1...6>:FUNCtion:DEModulation[:STATe]
<State>
This command turns the audio demodulator on and off when the measurement hits a
marker position.
With span > 0, you can define a hold time at the marker position with
CALCulate<1...2>:MARKer<1...6>:FUNCtion:DEModulation:HOLD
on page 320.
In zero span, the demodulation is on permanently.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Suffix:
<1...2>
.
1...2
<1...6>
1...6
Parameters:
<State>
ON | OFF
*RST:
OFF
Remote Control - Commands