Remote Commands
R&S
®
Spectrum Rider FPH
287
User Manual 1321.1011.02 ─ 06.00
*RST
Reset
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
The command is equivalent to
SYSTem:PRESet
.
Usage:
Setting only
*SRE
<Contents>
Service request enable
Sets the service request enable register to the indicated value. This command deter-
mines under which conditions a service request is triggered.
Parameters:
<Contents>
Contents of the service request enable register in decimal form.
Bit 6 (MSS mask bit) is always 0.
Range:
0 to 255
*STB?
Status byte query
Reads the contents of the status byte in decimal form.
Usage:
Query only
*TRG
Trigger
Triggers all actions waiting for a trigger event. In particular, *TRG generates a manual
trigger signal. This common command complements the commands of the
TRIGger
subsystem.
*TRG
corresponds to the
INITiate:IMMediate
command.
Usage:
Event
*TST?
Self-test query
Initiates self-tests of the instrument and returns an error code
Note:
If you start a self-test remotely, then select the "Local" softkey while the test is
still running, the instrument only returns to the manual operation state after the test is
completed. In this case, the self-test cannot be aborted.
Remote Control - Commands