Instrument Functions
R&S
®
Spectrum Rider FPH
104
User Manual 1321.1011.02 ─ 06.00
1. Press the [SETUP] key.
2. Select the "User Preference" softkey.
3. Select the "Capture Dataset" item and set it "On".
If the "Capture Dataset" menu item is set "On", the R&S Spectrum Rider saves a
dataset of the current measurement if an event occurs.
For more information on events, see
Chapter 4.7, "Saving On Events"
4.8.1.2
Renaming Filename
If necessary, you can rename files or file directories directly on the R&S Spectrum
Rider.
1. Press the SAVE/RECALL key.
2. Select the "File Manager" softkey.
The R&S Spectrum Rider opens the file manager.
3. Select the files or file directories in the file manager.
Alternatively, press the "Mark" softkey to select the desired files or files directories.
4. Select the "Edit" softkey.
The R&S Spectrum Rider displays a submenu for selection.
5. Select the "Rename" menu item to rename the files or file directories.
The R&S Spectrum Rider opens an input field to change the name of the selected
files or file directories.
Managing Datasets