Functions
R&S
®
OSP
205
User Manual 1178.7100.02 ─ 05
●
In active trigger state, the dialog shows the trigger events counter:
●
Also, this section displays the
"Minimum Trigger Interval"
in units of seconds,
which is derived from the data-sheet specifications of the modules and relays used
in the configured paths. The minimum interval comprises only the time that is
required to reach 90% of the target level of the switched state. It does not include
any operational time to be used while remaining in this state. However, the value
also considers a delay of approximately 1
µ
s for modules located in a satellite unit.
●
The
"Maximum Trigger Frequency"
in units of Hertz is the inverse value of the
"Minimum Trigger Interval".
Remote command:
TRIGger:STATe
TRIGger:EXECute
TRIGger:COUNt[:VALue]?
TRIGger:COUNt:OVERflow?
TRIGger:INDex?
Trigger Type
The "Trigger Type" section (middle portion of
) is accessible only, if the trig-
ger is
. It provides different functions and dialogs depending on the trig-
ger type.
Table 5-2: Triggerable paths per trigger type
Dialog in the user interface
Trigger type
Number of triggerable paths
1
2
3 to 16 ("Seq. Length")
3 to 16 ("Address Count")
The following illustration provides an overview of the different trigger functionalities
available with the different trigger types:
Configuration