Troubleshooting
R&S
®
NRPM
204
Mannual 1425.8663.02 ─ 08
Do not apply a signal to the sensor module while the selftest is running.
Executing the self-test with an active signal assigned, can erroneously result
in error messages at the test steps
Offset Voltages
and/or
Noise Voltages
.
To execute the selftest over remote control
► For a quick check, send
.
For each test step, PASS or FAIL is listed.
To execute the self test in the web browser-based user interface
(R&S
NRPM3N LAN sensor module)
1. In the navigation pane of the main dialog, select "Sensor".
2. Under "Diagnostics", select "Selftest".
See also
12.3
Problems during a firmware update
The firmware update is described in
Solutions for potential problems that can occur when using the firmware update
for R&S NRP family, see
"Firmware update for R&S NRP family"
Firmware update was interrupted
If, for example, a power cut happened during the firmware update, problems can
occur.
1. Perform the firmware update again. Sometimes, a further update fixes the
problems.
2. If you nevertheless can not access the sensor module, contact the local ser-
vice representative.
Problems during a firmware update