Preparing for use
R&S
®
NRPM
19
Mannual 1425.8663.02 ─ 08
EMI impact on measurement results
Electromagnetic interference (EMI) can affect the measurement results.
To suppress electromagnetic radiation during operation:
●
Use high-quality shielded cables, for example, double-shielded RF and inter-
face cables.
●
Always terminate open cable ends.
●
Ensure that connected external devices comply with EMC regulations.
●
When using the filtered cable feedthrough module R&S
NRPM-ZD3, make
through module is grounded by the housing.
Since the antenna module receives signals in a wide band, also outside the
specified frequency bands, always measure in a controlled EM environment.
In particular, when working with the R&S
NRPM-Z3 interface module.
Signal input and output levels
Information on signal levels is provided in the data sheet. Keep the signal levels
within the specified ranges to avoid damage to the product and connected devi-
ces.
Considerations for test setup