Remote control commands
R&S
®
NRPM
152
Mannual 1425.8663.02 ─ 08
Parameters:
<mode>
*RST:
"PISotropic"
Example:
CALC:MATH "PDEN"
Selects power density in W/m
2
as the equivalent resulting
unit.
CALCulate:MATH[:EXPRession]:CATalog?
Lists all supported calculation functions.
The result is a list of strings, separated by comma.
Example:
CALC:MATH:CAT?
Response:
"PISotropic","PDENsitiy",
"EFDensity","MFDensity","DETPower"
Usage:
Query only
[SENSe<Sensor>:]AUXiliary
<mode>
Activates the measurement of additional measurands that are determined
together with the main measured value.
Parameters:
<mode>
NONE | MINMax | RNDMax
NONE
No additional values are measured.
MINMax
In addition to the average trace, the sensor module also
determines the minimum and maximum traces. You can
read out the resulting traces (min & max) afterwards with
the command
.
RNDMax
Determines a trace of randomly selected samples within
the min & max range instead of the minimum trace. You
can read out the resulting traces (random & max) after-
wards with the command
.
*RST:
NONE
Selecting a measurement mode and retrieving results