Remote control commands
R&S
®
NRPM
141
Mannual 1425.8663.02 ─ 08
p
t
A
B
C
D
E
F
G
H
1
2
3
1 = Hold-off time
2 = Trigger hysteresis
3 = Trigger level
10.5.2
Controlling the measurement results
The R&S
NRPM3(N) sensor module can cope with the wide range of measure-
ment scenarios with the help of the so-called "termination control". Depending on
how fast your measurement results change, you can define, how to retrieve the
measurement results.
In continuous average mode, use
[SENSe<Sensor>:]AVERage:TCONtrol
[SENSe<Sensor>:]TRACe:AVERage:TCONtrol
.
Repeating termination control
Returns a measurement result when the average filter has been filled completely.
The number of measurement cycle repetitions is equal to the set average count. If
the average count is large, the measurement time can take long time.
Useful if you expect slow changes in the results, and you want to avoid the output
of redundant data.
Select repeating termination control when executing remote control applica-
tions, as you are usually only interested in the results of the fully settled
average filter.
Controlling the measurement