Remote Control - Commands
NRP2
294
User Manual 1173.9140.02 ─ 07
SERVice<[1]..4>:CALibration:TEST?
<NR1>
Tests the consistency of the calibration data set according to three criteria:
If the following bit is set in
<nr1>,
the following action is executed.
0
The serial number of the calibration data set is tested.
1
The checksum of the calibration data set is tested.
2
The header of the calibration data set is tested.
If an inconsistency of the data set is detected, the query returns a 1, otherwise a 0.
Parameters:
<NR1>
int_value
Range:
0 to 255
Usage:
Query only
SERVice<[1]..4>:CALibration:TEMP
<state>
Initiates a temperature measurement of the sensor in question. The temperature can
be queried with
SERV:CAL:TEMP:DATA?
.
Parameters:
<state>
ONCE
Usage:
Event
SERVice<[1]..4>:CALibration:TEMP:DATA?
Returns the temperature of the sensor in Kelvin if a temperature measurement was
previously initiated by means of
SERV:CAL:TEMP ONCE
.
Usage:
Query only
SERVice:DISPlay:OVERload
<state>
Switches the sensor overload warning on/off.
Attention:
If the sensor is overloaded, it could be damaged! The warning should not be sup-
pressed if you are not certain that the sensor is out of danger.
Parameters:
<state>
ON | OFF
*RST:
ON
SERVice