Remote Control - Commands
NRP2
281
User Manual 1173.9140.02 ─ 07
[SENSe<[1]..4>]:STATistics:PEAK?
Returns the peak power value calculated during a statistics measurement (CCDF/
PDF).
Usage:
Query only
[SENSe<[1]..4>]:STATistics:OFFSet[:TIME]
<time_interval>
Determines, together with the
SENS:STAT:TIME
command, the time interval in which
the power for the statistical evaluation is measured.
SENS:STAT:OFFS
determines the
start of the interval relative to the (possibly delayed) trigger time (see also
[SENSe<[1]..4>][:POWer]:TGATe<[1]..4>:OFFSet:TIME
Parameters:
<time_interval>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: S
[SENSe<[1]..4>]:STATistics:SAMPles[:MINimum]
<minimum>
Determines the minimum number of samples to be included in the statistics. For this
purpose, the filter length (which can only be set in powers of 2) is set such that the fol-
lowing inequality applies:
dt
mpw
N
N
MinSamples
fil
In this inequality,
N
fil
is the filter length (
SENS:TRAC:AVER:COUN
),
mpw
the time width
of a sample and dt the length of the window in which the measurement is performed
(
SENS:STAT:TIME
).
Parameters:
<minimum>
int_value
*RST:
1e6
[SENSe<[1]..4>]:STATistics:SCALe:X:MPWidth?
Returns the minimum width of a sample on the power axis.
Parameters:
<mpwidth>
float_value
Usage:
Query only
SENSe (Sensor Commands)