SCPI Commands HMF series
Remote Control
18
Command Reference
SCPI Programmers Manual
SYSTem:SNUM?
Queries the serial number of the instrument.
Usage:
Query only
SYSTem:TREE?
Returns a list of implemented remote commands.
Usage:
Query only
SYSTem:ELISt?
Queries the error list.
Usage:
Query only
SYSTem:ERRor[:NEXT]?
Queries an error and removes it from the queue. Positive error numbers are instrument-
dependent. Negative error numbers are reserved by the SCPI standard. If the queue is empty the
response is 0, “No error“.
Return values:
<error>
0
, “No error“
-100
, “Command error“
-102
, “Syntax error“
-109
, “Missing parameter
-200
, “Execution error“
-350
, “Queue overflow“
Usage:
Query only
SYSTem:ERRor:ALL? <Error>
Queries the error/event queue for all unread items and removes them from the queue.
The response is a comma separated list of error number and a short description of the error in
FIFO order. Positive error numbers are instrument-dependent. Negative error numbers are
reserved by the SCPI standard.
Return values:
<Error>
List of: Error/event_number,“Error/event_description>[;Devicedependent info]“
If the queue is empty, the response is 0,“No error”
Usage:
Query only
SCPI confirmed