SCPI Commands HMF series
Remote Control
15
Command Reference
SCPI Programmers Manual
NOTICE
We recommend to start a program by *RST in order to set the instrument to a defined
status prior to starting a program.
*SRE <Contents>
Service Request Enable
Sets the service request enable register to the indicated value. This command determines under
which conditions a service request is triggered. The query *SRE? returns a decimal value of the
service request enable register which corresponds to the binary-weighted sum of all bits.
Parameters:
<Contents>
Contents of the service request enable register in decimal form.
Bit 6 (MSS mask bit) is always 0.
Range:
0 to 255
NOTICE
The SRE is an enable register. Consequently, there are no denotations about the bits.
This register conduce for the „OR“ combination of the bits in the status byte.
*STB?
STatus Byte query
Returns the contents of the status byte in decimal form.
Usage:
Query only
*WAI
WAIt to continue
Prevents servicing of the subsequent commands until all preceding commands have been exe-
cuted.
Usage:
Event