Remote Commands
R&S
®
FSW
993
User Manual 1173.9411.02 ─ 19
MMEMory:CDIRectory
<Directory>
This command changes the current directory.
Parameters:
<Directory>
String containing the path to another directory.
The path may be relative or absolute.
Usage:
SCPI confirmed
MMEMory:COMMent
<Comment>
This command defines a comment for the stored settings.
Parameters:
<Comment>
String containing the comment.
Example:
MMEMory:COMMent "ACP measurement with Standard
Tetra from 23.05."
MMEMory::MMEMory:STORe1:STATe 1, "ACP_T"
As a result, in the selection list for recall settings, the comment
"ACP measurement with Standard Tetra from
23.05."
is added to the ACP entry.
Manual operation:
See
MMEMory:COPY
<SourceFile>,<DestinationFile>
This command copies one or more files to another directory.
Parameters:
<SourceFile>
String containing the path and file name of the source file.
<DestinationFile>
String containing the path and name of the target file.
The path may be relative or absolute.
Usage:
SCPI confirmed
MMEMory:DATA
<FileName>, [<Block>]
This command writes block data into a file. The delimiter must be set to EOI to obtain
error-free data transfer.
When you query the contents of a file, you can save them in a file on the remote con-
trol computer.
The command is useful for reading stored settings files or trace data from the instru-
ment or for transferring them to the instrument.
Parameters:
<FileName>
String containing the path and name of the target file.
Managing Settings and Results