Remote Commands
R&S
®
FSW
964
User Manual 1173.9411.02 ─ 19
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 1.
CALC:MARK:COUN ON
Switches on the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value.
Manual operation:
See
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
This command queries the frequency at the marker position.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also
Before you can use the command, you have to turn on the frequency counter.
Return values:
<Frequency>
Frequency at the marker position.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK ON
Switches on marker 2.
CALC:MARK:COUN ON
Activates the frequency counter for marker 1.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:COUN:FREQ?
Outputs the measured value of marker 1.
Usage:
Query only
Manual operation:
See
CALCulate<n>:MARKer<m>:COUNt:RESolution
<Resolution>
This command defines the resolution of the frequency counter.
Parameters:
<Resolution>
0.001 | 0.01 | 0.1 | 1 | 10 | 100 | 1000 | 10000 Hz
*RST:
0.1 Hz
Example:
CALC:MARK:COUN:RES 1kHz
Sets the resolution of the frequency counter to 1 kHz.
Manual operation:
See
Analyzing Measurements (Basics)