Remote Commands
R&S
®
FSW
959
User Manual 1173.9411.02 ─ 19
Return values:
<Power>
Signal power over the marker bandwidth.
Example:
Activate the band power marker:
CALC:MARK:FUNC:BPOW:STAT ON
Select the density mode for the result:
CALC:MARK:FUNC:BPOW:MODE DENS
Query the result:
CALC:MARK:FUNC:BPOW:RES?
Response:
20dBm/Hz
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:SPAN
<Span>
This command defines the bandwidth around the marker position.
Parameters:
<Span>
Frequency. The maximum span depends on the marker position
and R&S
FSW model.
*RST:
5% of current span
Default unit: Hz
Example:
CALC:MARK:FUNC:BPOW:SPAN 2MHz
Measures the band power over 2 MHz around the marker.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer[:STATe]
<State>
This command turns markers for band power measurements on and off.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MARK4:FUNC:BPOW:STAT ON
Activates or turns marker 4 into a band power marker.
Manual operation:
See
"Band Power Measurement State"
See
"Switching All Band Power Measurements Off"
Using Delta Markers
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:MODE
............................................ 960
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:RESult?
..........................................960
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:SPAN
............................................. 960
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer[:STATe]
..........................................960
Analyzing Measurements (Basics)