![Rohde & Schwarz FSW13 User Manual Download Page 943](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847943.webp)
Remote Commands
R&S
®
FSW
943
User Manual 1173.9411.02 ─ 19
Usage:
Query only
Manual operation:
See
See
CALCulate<n>:DELTamarker<m>:Y?
This command queries the relative position of a delta marker on the y-axis.
If necessary, the command activates the delta marker first.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also
The unit depends on the application of the command.
Table 11-9: Base unit
Parameter, measuring function or result display
Output unit
DBM | DBPW | DBUV | DBMV | DBUA
dB (lin/log)
WATT | VOLT | AMPere
dB (lin), % (log)
statistics function (APD or CCDF) on
dimensionless output
Return values:
<Position>
Position of the delta marker in relation to the reference marker or
the fixed reference.
Example:
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
CALC:DELT2 ON
Switches on delta marker 2.
CALC:DELT2:Y?
Outputs measurement value of delta marker 2.
Usage:
Query only
CALCulate<n>:MARKer<m>:Y?
This command queries the position of a marker on the y-axis.
If necessary, the command activates the marker first.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also
Return values:
<Result>
Result at the marker position.
The unit is variable and depends on the one you have currently
set.
Analyzing Measurements (Basics)