Remote Commands
R&S
®
FSW
924
User Manual 1173.9411.02 ─ 19
Table 11-7: Return values for LIST parameter
For each peak, the command returns 11 values in the following order:
<No>,<StartFreq>,<StopFreq>,<RBW>,<PeakFreq>,<PowerAbs>,<PowerRel>,<PowerDelta>,<Limit-
Check>,<Unused1>,<Unused2>
●
<No>: range number
●
<StartFreq>,<StopFreq>: start and stop frequency of the range
●
<RBW>: resolution bandwidth
●
<PeakFreq>: frequency of the peak in a range
●
<PowerAbs>: absolute power of the peak in dBm
●
<PowerRel>: power of the peak in relation to the channel power in dBc
●
<PowerDelta>: distance from the peak to the limit line in dB, positive values indicate a failed limit
check
●
<LimitCheck>: state of the limit check (0 = PASS, 1 = FAIL)
●
<Unused1>,<Unused2>: reserved (0.0)
Table 11-8: Return values for SPECtrogram parameter
For every frame in the spectrogram, the command returns the power levels that have been measured, one
for each sweep point. The number of frames depends on the size of the history depth. The power level
depends on the unit you have currently set.
TRACe<n>[:DATA]:MEMory?
<Trace>,<OffsSwPoint>,<NoOfSwPoints>
This command queries the previously captured trace data for the specified trace from
the memory. As an offset and number of sweep points to be retrieved can be specified,
the trace data can be retrieved in smaller portions, making the command faster than
the
TRAC:DATA?
command. This is useful if only specific parts of the trace data are of
interest.
If no parameters are specified with the command, the entire trace data is retrieved; in
this case, the command is identical to
TRAC:DATA? TRACE1
For details on the returned values see the
command.
Query parameters:
<Trace>
TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6
<OffsSwPoint>
The offset in sweep points related to the start of the measure-
ment at which data retrieval is to start.
<NoOfSwPoints>
Number of sweep points to be retrieved from the trace.
Example:
TRAC:DATA:MEM? TRACE1,25,100
Retrieves 100 sweep points from trace 1, starting at sweep point
25.
Usage:
Query only
TRACe<n>[:DATA]:X?
<TraceNumber>
This command queries the horizontal trace data for each sweep point in the specified
window, for example the frequency in frequency domain or the time in time domain
measurements.
This is especially useful for traces with non-equidistant x-values, e.g. for SEM or Spuri-
ous Emissions measurements.
Analyzing Measurements (Basics)