Remote Commands
R&S
®
FSW
877
User Manual 1173.9411.02 ─ 19
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:DATA 1MHZ,-30DB,2MHZ,-40DB
Manual operation:
See
[SENSe:]CORRection:CVL:HARMonic
<HarmOrder>
This command defines the harmonic order for which the conversion loss table is to be
used. This setting is checked against the current mixer setting before the table can be
assigned to the range.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<HarmOrder>
numeric value
Range:
2 to 65
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:HARM 3
Manual operation:
See
[SENSe:]CORRection:CVL:MIXer
<Type>
This command defines the mixer name in the conversion loss table. This setting is
checked against the current mixer setting before the table can be assigned to the
range.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<Type>
string
Name of mixer with a maximum of 16 characters
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:MIX 'FS_Z60'
Manual operation:
See
[SENSe:]CORRection:CVL:PORTs
<PortNo>
This command defines the mixer type in the conversion loss table. This setting is
checked against the current mixer setting before the table can be assigned to the
range.
Setting Basic Measurement Parameters