Remote Commands
R&S
®
FSW
863
User Manual 1173.9411.02 ─ 19
Parameters:
<Threshold>
Range:
0 dB to 200 dB
*RST:
+1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:UPP 2
Example:
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level rises above 22
dBm.
Manual operation:
See
[SENSe:]ADJust:CONFigure:TRIG
<State>
Defines the behaviour of the measurement when adjusting a setting automatically
(using
SENS:ADJ:LEV ON
, for example).
See
"Adjusting settings automatically during triggered measurements"
Parameters:
<State>
ON | 1
The measurement for automatic adjustment waits for the trigger.
OFF | 0
The measurement for automatic adjustment is performed imme-
diately, without waiting for a trigger.
*RST:
1
[SENSe:]ADJust:FREQuency
This command sets the center frequency to the highest signal level in the current fre-
quency range.
Example:
ADJ:FREQ
Usage:
Event
Manual operation:
See
"Adjusting the Center Frequency Automatically (Auto Freq)"
[SENSe:]ADJust:LEVel
This command initiates a single (internal) measurement that evaluates and sets the
ideal reference level for the current input data and measurement settings. This ensures
that the settings of the RF attenuation and the reference level are optimally adjusted to
the signal level without overloading the R&S
FSW or limiting the dynamic range by an
S/N ratio that is too small.
Example:
ADJ:LEV
Usage:
Event
Manual operation:
See
"Setting the Reference Level Automatically (Auto Level)"
Setting Basic Measurement Parameters