Remote Commands
R&S
®
FSW
809
User Manual 1173.9411.02 ─ 19
CALCulate:DELTamarker<m>:FUNCtion:FMEasurement:LIMit<k>:DELTa?
This command queries the vertical distance from the deltamarker position to the limit
line. The unit is dB.
If the deltamarker has been assigned to a different trace than the limit line, or if no limit
ine is defined for the marker position, the command returns -200.
Return values:
<Amplitude>
Vertical distance to the limit line in dB.
Example:
CALC:DELT3:FUNC:FME:LIM2:DELT?
Queries the distance of deltamarker 3 to the second limit line.
Usage:
Query only
CALCulate:MARKer<m>:FUNCtion:FMEasurement:LIMit<k>:DELTa?
<Amplitude>
This command queries the vertical distance from the marker position to the limit line.
The unit is dB.
If the marker has been assigned to a different trace than the limit line, or if no limit ine
is defined for the marker position, the command returns -200.
Return values:
<Amplitude>
Vertical distance to the limit line in dB.
Example:
CALC:MARK3:FUNC:FME:LIM2:DELT?
Queries the distance of marker 3 to the second limit line.
Usage:
Query only
11.5.13.7
Evaluating the Results
The commands required to control the demodulation of signals at the marker position
are described in
chapter 11.8.3.14, "Marker Demodulation"
11.5.13.8
Programming Example: EMI Measurement
This example demonstrates how to detect electromagnetic interferences (EMI) in a
remote environment.
//----------- Preparing the measurement ------------
//Reset the instrument
*RST
//Define the span to be analyzed
FREQ:STAR 150kHz
FREQ:STOP 1GHz
//Configure two traces, one with peak detector, one with average detector
DISP:TRAC1 ON
DISP:TRAC2 ON
DET1 POS
DET2 AVER
Configuring and Performing Measurements