Remote Commands
R&S
®
FSW
716
User Manual 1173.9411.02 ─ 19
[SENSe:]POWer:ACHannel:BANDwidth|BWIDth:GAP<gap>
<Bandwidth>
This command defines the bandwidth of the specified MSR gap (CACLR) channel.
This command is for MSR signals only (see
Suffix:
<gap>
.
1 | 2
Gap (CACLR) channel number
Parameters:
<Bandwidth>
numeric value in Hz
*RST:
3.84 MHz
Manual operation:
See
"Gap (CACLR) Channel Bandwidths"
[SENSe:]POWer:ACHannel:FILTer:ALPHa:GAP<gap>
<Alpha>
This command defines the roll-off factor for the specified gap (CACLR) channel's
weighting filter.
Suffix:
<gap>
.
1 | 2
Gap (CACLR) channel number
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Manual operation:
See
[SENSe:]POWer:ACHannel:FILTer:ALPHa:SBLock<sb>:CHANnel<ch>
<Alpha>
This command defines the roll-off factor for the specified transmission channel's
weighting filter.
Suffix:
<sb>
.
1 | 2 | 3 | 4 | 5
sub block number
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Manual operation:
See
[SENSe:]POWer:ACHannel:FILTer:STATe:GAP<gap>
<State>
This command turns the weighting filter for the specified gap (CACLR) channel on and
off.
Configuring and Performing Measurements