Remote Commands
R&S
®
FSW
710
User Manual 1173.9411.02 ─ 19
You have to activate the general ACLR limit check before using this command with
.
Parameters:
<State>
ON | OFF
*RST:
OFF
Manual operation:
See
See
CALCulate:LIMit:ACPower:ALTernate<ch>:ABSolute
<LowerLimit>, <UpperLimit>
This command defines the absolute limit of the alternate channels.
If you have defined an absolute limit as well as a relative limit, the R&S
FSW uses the
lower value for the limit check.
Parameters:
<LowerLimit>,
<UpperLimit>
The first value defines the limit of the lower alternate channel,
the second value the limit of the upper alternate channel.
Range:
-200 dBm to 200 dBm
*RST:
-200 dBm
Manual operation:
See
See
CALCulate:LIMit:ACPower:ALTernate<ch>:ABSolute:STATe
<State>
This command turns the absolute limit check for the alternate channels on and off.
You have to activate the general ACLR limit check before using this command with
.
Parameters:
<State>
ON | OFF
*RST:
OFF
Manual operation:
See
See
CALCulate:LIMit:ACPower:ALTernate<ch>[:RELative]
<LowerLimit>, <UpperLimit>
This command defines the relative limit of the alternate channels. The reference value
for the relative limit is the measured channel power.
If you have defined an absolute limit as well as a relative limit, the R&S
FSW uses the
lower value for the limit check.
Configuring and Performing Measurements