Common Analysis and Display Functions
R&S
®
FSW
520
User Manual 1173.9411.02 ─ 19
2. If the required signal point is not the maximum, continue the peak search to one of
the subsequent maxima or minima:
a) Press the MKR -> key.
b) Select the "Next Peak" or "Next Min" key.
c) If necessary, change the search settings by selecting the "Search Config" soft-
key.
3. Center the display around the determined signal point by setting the marker value
to the center frequency. Select the "Center = Mkr Freq" softkey.
4. Determine the precise frequency of the signal point:
a) Select the "Select Marker Function" softkey.
b) Select the "Signal Count" button.
c) Select the "Signal Count Resolution" softkey.
d) Select the resolution depending on how precise the result needs to be.
7.4.4.2
How to Use a Fixed Reference Marker
By default, delta markers refer to marker 1. However, they can also refer to a fixed ref-
erence marker.
How to Define and Move a Fixed Reference Marker
1. To display a fixed reference marker, do one of the following:
● Press the MKR FUNC key, then select the "Reference Fixed" marker function.
● In the "Marker" dialog box, in the "Reference Fixed" area of the "Marker Config"
tab, set the "State" to "On".
A vertical and a horizontal red display line are displayed, marked as "FXD". The
normal marker 1 is activated and set to the peak value of the trace assigned to
marker 1, and a delta marker to the next peak. The fixed reference marker is set to
the position of marker 1 at the peak value.
2. To move the fixed reference marker, do one of the following:
● Change the "Level" and "Frequency" of the reference point in the "Marker Con-
fig" tab of the "Marker" dialog box, . By default, the current peak value of trace
1 is set.
● Set the fixed reference marker to the current peak value by selecting the "Peak
Search" button in the "Marker Config" tab of the "Marker" dialog box.
● Move the "FXD" display lines that define the position of the fixed reference
marker by dragging them on the screen.
How to Assign a Fixed Reference Marker to Delta Markers
1. In the "Marker" dialog box, select the horizontal "Markers" tab.
Marker Usage