Common Measurement Settings
R&S
®
FSW
416
User Manual 1173.9411.02 ─ 19
This function is only available in single sweep mode or if the sweep is stopped, and
only if a spectrogram is selected.
The most recent frame is number 0, all previous frames have a negative number.
For more information see
Remote command:
Continue Frame ← Spectrogram Frames
Determines whether the results of the previous sweeps are included in the analysis of
the next sweeps for trace modes "Max Hold", "Min Hold", and "Average".
This function is available in single sweep mode only.
●
On
When the average or peak values are determined for the new sweep, the results of
the previous sweeps in the spectrogram are also taken into account.
●
Off
The average or peak values are determined from the results of the newly swept
frames only.
Remote command:
Frame Count ← Spectrogram Frames
Determines how many frames are plotted during a single sweep measurement (as
opposed to a continuous sweep). The maximum number of possible frames depends
on the history depth (see
For more details see
Remote command:
Clear Spectrogram ← Spectrogram Frames
Resets the spectrogram result display and clears the history buffer.
This function is only available if a spectrogram is selected.
Remote command:
6.5.3 Reference: List of Available RRC and Channel Filters
For power measurement a number of especially steep-edged channel filters are availa-
ble (see the following table). The indicated filter bandwidth is the 3 dB bandwidth. For
RRC filters, the fixed roll-off factor (a) is also indicated.
The available Gaussian 3dB sweep filters are listed in the R&S
FSW data sheet.
Bandwidth, Filter and Sweep Configuration