Common Measurement Settings
R&S
®
FSW
414
User Manual 1173.9411.02 ─ 19
For zero span measurements, the optimization mode defines the selection of the A/D
converter prefilter.
Table 6-3: Optimization parameters in zero span mode
Optimization mode
Description
Dynamic
The narrowest filter possible (depending on the RBW) is used.
Speed
The widest filter possible (depending on the RBW) is used.
Auto
A medium-sized prefilter is used.
Note: EMI measurements
For EMI measurements (using R&S
FSW-K54), "Dynamic" mode is not supported.
"Auto" mode always uses "Speed" optimization.
Remote command:
Sweep Type
Defines the sweep type.
"Auto"
Automatically sets the fastest available sweep type for the current
measurement (Frequency or FFT). Auto mode is set by default.
"FFT"
The FFT sweep samples on a defined frequency value and trans-
forms it to the spectrum by fast Fourier transformation (FFT) (see
also
chapter 6.5.1.5, "How Data is Measured: the Sweep Type"
on page 405).
FFT is not available in the following cases:
●
when using 5-Pole filters, Channel filters or RRC filters
●
when using the Quasi peak detector
●
when an external generator (option B10) is active
In these cases, frequency sweep is used.
Remote command:
Single Sweep/ RUN SINGLE
After triggering, starts the number of sweeps set in "Sweep Count". The measurement
stops after the defined number of sweeps has been performed.
While the measurement is running, the "Single Sweep" softkey and the RUN SINGLE
key are highlighted. The running measurement can be aborted by selecting the high-
lighted softkey or key again.
Note:
Sequencer. If the Sequencer is active, the "Single Sweep" softkey only controls
the sweep mode for the currently selected channel; however, the sweep mode only
has an effect the next time the Sequencer activates that channel, and only for a chan-
nel-defined sequence. In this case, a channel in single sweep mode is swept only once
by the Sequencer.
Furthermore, the RUN SINGLE key controls the Sequencer, not individual sweeps.
RUN SINGLE starts the Sequencer in single mode.