Common Measurement Settings
R&S
®
FSW
412
User Manual 1173.9411.02 ─ 19
"Manual"
The coupling ratio is defined manually.
The RBW/VBW ratio can be set in the range of 0.001 to 1000.
Remote command:
Filter Type
Defines the filter type.
The following filter types are available:
●
Normal (3dB)
●
Channel
●
RRC
●
5-Pole (not available for sweep type "FFT")
●
CISPR (6 dB) - requires EMI (R&S
FSW-K54) option
●
MIL Std (6 dB) - requires EMI (R&S
FSW-K54) option
For more information see
chapter 6.5.1.6, "Which Data May Pass: Filter Types"
Note:
The EMI-specific filter types are available if the EMI (R&S
FSW-K54) measure-
ment option is installed, even if EMI measurement is not active. For details see
ter 5.13.3.1, "Resolution Bandwidth and Filter Types"
The RBW filter configured in the bandwidth settings is identical to the filter configured
in the EMI configuration.
Remote command:
Default Coupling
Sets all coupled functions to the default state ("AUTO"). In addition, the ratio "RBW/
VBW" is set to "SINE [1/1]" and the ratio "SPAN/RBW" to
100.
For more information see
chapter 6.5.1.3, "Coupling VBW and RBW"
Remote command:
Sweep/Average Count
Defines the number of sweeps to be performed in the single sweep mode. Values from
0 to 200000 are allowed. If the values 0 or 1 are set, one sweep is performed.
The sweep count is applied to all the traces in all diagrams.
If the trace modes "Average", "Max Hold" or "Min Hold" are set, this value also deter-
mines the number of averaging or maximum search procedures.
In continuous sweep mode, if sweep count = 0 (default), averaging is performed over
10 sweeps. For sweep count =1, no averaging, maxhold or minhold operations are per-
formed.
For more information see
chapter 6.5.1.8, "How Much Data is Measured: Sweep Points
Bandwidth, Filter and Sweep Configuration