Measurements
R&S
®
FSW
214
User Manual 1173.9411.02 ─ 19
For more information see
"Relative limit line functions"
Remote command:
Multi-Limit Calc <n>
Defines the function used to calculate the limit line for the n-th power class for overlap-
ping ranges in Multi-SEM measurements. For details see
"Limit calculation for individ-
"NONE"
(reference ranges only:) the limit of the reference range is used
"SUM"
sum of the two limit lines (calculated for linear powers) is used
"MAX"
maximum of the two limit lines is used
Remote command:
Minimum Sweep Points
Defines the minimum number of sweep points for the range.
If necessary to fulfill all minimum sweep point requirements in all ranges, the global
setting is increased. By default, each range is supplied with a minimum
of one sweep point.
For details see
Remote command:
Insert before/after Range
Inserts a new range to the left (before) or to the right (after) of the range in which the
cursor is currently displayed. The range numbers of the currently focused range and all
higher ranges are increased accordingly. The maximum number of ranges is 30.
Remote command:
Spectrum Emission Mask (SEM) Measurement