Measurements
R&S
®
FSW
197
User Manual 1173.9411.02 ─ 19
on page 224). By default, one peak per range is displayed. However, you can change
the settings to display only peaks that exceed a threshold ("Margin").
In addition to listing the peaks in the Result Summary, detected peaks can be indicated
by blue squares in the diagram (see
in the "List Evaluation" settings).
Fig. 5-31: Detected peak display in SEM measurement
Furthermore, you can export the results of the result summary to a file which can be
exported to another application for further analysis.
Results for SEM with multiple sub blocks
As of firmware version 1.91, and in the Spectrum application only, spectrum emissions
can be measured for multiple sub blocks of channels (see
Multiple Sub Blocks ("Multi-SEM")"
on page 205 for details). Up to three sub blocks
(with two gaps) can be defined. For each sub block and each gap, the results descri-
bed above are provided individually in the result summary.
Spectrum Emission Mask (SEM) Measurement