Measurements
R&S
®
FSW
196
User Manual 1173.9411.02 ─ 19
Fig. 5-30: Spectrum Emission Mask result displays
In addition to the graphical results of the SEM measurement displayed in the diagram,
a result summary is displayed to evaluate the limit check results (see also
ter 5.6.4.2, "Limit Lines in SEM Measurements"
The following information is provided in the result summary:
Label
Description
General Information
Standard
Loaded standard settings
Tx Power
Power of the reference range
Tx Bandwidth
Tx bandwidth used by the reference range
RBW
RBW used by the reference range
Range results
Range Low
Frequency range start the peak value was found in
Range Up
Frequency range end the peak value was found in
RBW
RBW of the range
Frequency
Frequency of the peak power level
Power Abs
Absolute peak power level within the range
Power Rel
Peak power level within the range, relative to the "Tx Power"
Δ
Limit
Deviation of the peak power level from the limit line
You can define in which detail the data is displayed in the result summary in the "List
Evaluation" settings (see
chapter 5.6.5.7, "List Evaluation (Results Configuration)"
Spectrum Emission Mask (SEM) Measurement