Optimizing and Troubleshooting
R&S
®
FSW
1101
User Manual 1173.9411.02 ─ 19
The following keywords are used:
DATA ERR
For Digital Baseband Interface (R&S
FSW-B17) only:
Error in digital I/Q input data
For details on the Digital Baseband Interface (R&S
FSW-B17) see the R&S
FSW I/Q
Analyzer User Manual.
FIFO OVLD
For Digital Baseband Interface (R&S
FSW-B17) only:
Input sample rate from connected instrument is too high
For details on the Digital Baseband Interface (R&S
FSW-B17) see the R&S
FSW I/Q
Analyzer User Manual.
IF OVLD
Overload of the IF signal path after the input mixer.
●
Increase the reference level.
INPUT OVLD
The signal level at the RF input connector exceeds the maximum.
The RF input is disconnected from the input mixer to protect the device. In order to
re-enable measurement, decrease the level at the RF input connector and reconnect
the RF input to the mixer input.
LOUNL
Error in the instrument's frequency processing hardware was detected.
NO REF
Instrument was set to an external reference but no signal was detected on the refer-
ence input.
OVEN
OCXO reference frequency (option R&S
FSW-B4) has not yet reached its operating
temperature. The message usually disappears a few minutes after power has been
switched on.
OVLD
Overload of the input signal path after the input mixer (only when RF INPUT path is
NOT used, e.g. for input from the Digital Baseband Interface (R&S
FSW-B17) or the
Analog Baseband Interface (R&S
FSW-B71)).
●
Reduce the input level.
PLL UNLOCK
For Digital Baseband Interface (R&S
FSW-B17) only:
Error in digital I/Q input data
For details on the Digital Baseband Interface (R&S
FSW-B17) see the R&S
FSW I/Q
Analyzer User Manual.
RF OVLD
Overload of the input mixer.
●
Increase the RF attenuation (for RF input).
●
Reduce the input level (for digital input)
UNCAL
One of the following conditions applies:
●
Correction data has been switched off.
●
No correction values are available, for example after a firmware update.
●
Record the correction data by performing a self alignment
(For details refer to
chapter 3.1.1.6, "Performing a Self-Alignment and a Selft-
WRONG_FW
The firmware version is out-of-date and does not support the currently installed hard-
ware. Until the firmware version is updated, this error message is displayed and self-
alignment fails.
(For details refer to
chapter 9.3.4.4, "Firmware Updates"
Error Information