Applications and Operating Modes
R&S
®
FSW
107
User Manual 1173.9411.02 ─ 19
Tab name
application
Sweep mode
Sweep count
Spectrum 3
Spectrum
Cont. Sweep
2
IQ Analyzer
IQ Analyzer
Single Sweep
7
For
single Sequence
, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer
For
continuous Sequence
, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
...
For
channel-defined Sequence
, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
5x Spectrum, 2 x Spectrum 3,
5x Spectrum, 2 x Spectrum 3,
...
RUN SINGLE/RUN CONT and Single Sweep/Sweep Continuous keys
While the Sequencer is active, the RUN SINGLE and RUN CONT keys control the
Sequencer, not individual sweeps. RUN SINGLE starts the Sequencer in single mode,
while RUN CONT starts the Sequencer in continuous mode.
The "Single Sweep" and "Continuous Sweep"
softkeys
control the sweep mode for the
currently selected channel only; the sweep mode only has an effect the next time the
Sequencer activates that channel, and only for a channel-defined sequence. In this
case, a channel in single sweep mode is swept only once by the Sequencer. A channel
in continuous sweep mode is swept repeatedly.
4.5.2 Sequencer Settings
The "Sequencer" menu is available from the toolbar.
Sequencer in MSRA/MSRT mode
In MSRA/MSRT operating mode and in the Realtime Spectrum Application, the
Sequencer behaves slightly differently.
For details on the MSRA operating mode see the R&S
FSW MSRA User Manual. For
details on the MSRT operating mode and the Realtime Spectrum application see the
R&S
FSW Realtime Spectrum Application and MSRT Operating Mode User Manual.
Running a Sequence of Measurements