Remote Commands
R&S
®
FSW
1004
User Manual 1173.9411.02 ─ 19
The file name depends on the one you have set with
. This
command adds a consecutive number to the file name.
Secure User Mode
In secure user mode, settings that are to be stored on the instrument are stored to vol-
atile memory, which is restricted to 256
MB. Thus, a "Memory full" error may occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details see
chapter 3.1.7, "Protecting Data Using the Secure User Mode"
For details see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSW User Manual.
Example:
MMEM:STOR:STAT 1,'Save'
Saves the current device settings in the file
Save.dfl
.
MMEM:STOR:STAT:NEXT
Saves the current device settings in the file
Save_001.dfl
MMEM:STOR:STAT:NEXT
Saves the current device settings in the file
Save_002.dfl
Usage:
Event
Manual operation:
See
MMEMory:STORe:TYPE
<Mode>
This command defines whether the data from the entire instrument or only from the
current channel is stored with the subsequent
MMEM:STOR
... command.
Parameters:
<Mode>
INSTrument | CHANnel
*RST:
INST
Example:
INST:SEL 'SPECTRUM2'
Selects measurement channel
'SPECTRUM2'
.
MMEM:STOR:TYPE CHAN
Specifies that channel-specific data is to be stored.
MMEM:STOR:PEAK 'SpectrumPeaks'
Stores the peak list from channel
'SPECTRUM2'
to the file
'SpectrumPeaks'
.
SYSTem:PRESet
This command presets the R&S
FSW.
Example:
SYST:PRES
Usage:
Event
SCPI confirmed
Managing Settings and Results