Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
290
User Manual 1173.9240.02 ─ 17
[SENSe:]ADEMod:AM[:ABSolute][:TDOMain]:RESult?
<TraceMode>
[SENSe:]ADEMod:AM:RELative[:TDOMain]:RESult?
<TraceMode>
[SENSe:]ADEMod:AM:RELative:AFSPectrum:RESult?
<TraceMode>
[SENSe:]ADEMod:FM[:TDOMain]:RESult?
<TraceMode>
[SENSe:]ADEMod:FM:AFSPectrum:RESult?
<TraceMode>
[SENSe:]ADEMod:PM[:TDOMain]:RESult?
<TraceMode>
[SENSe:]ADEMod:PM:AFSPectrum:RESult?
<TraceMode>
[SENSe:]ADEMod:SPECtrum:RESult?
<TraceMode>
This command reads the result data of the evaluated signal in the specified trace
mode. The data format of the output data block is defined by the FORMat command
(see
The trace results are configured for a specific evaluation. The following table indicates
which command syntax refers to which evaluation method, as well as the output unit of
the results.
Command syntax
Evaluation method
Output unit
AM[:ABSolute][:TDOMain]
RF time domain
dBm
AM:RELative[:TDOMain]
AM time domain
%
AM:RELative:AFSPectrum
AM spectrum
%
FM[:TDOMain]
FM time domain
kHz
FM:AFSPectrum
FM spectrum
kHz
PM[:TDOMain]
PM time domain
rad or °
PM:AFSPectrum
PM spectrum
rad or °
SPECtrum
RF spectrum
dBm (logarithmic display)
or V (linear display).
Query parameters:
<TraceMode>
WRITe | AVERage | MAXHold | MINHold | VIEW
The specified trace mode must be one of those configured by
SENS:ADEM:<Evaluation>:TYPE
, see
on page 273. Otherwise a query error is
generated.
Example:
ADEM:AM AVER,MAXH,MINH
Sets up RF time domain results to be measured
INIT; *WAI
Starts measurement and waits for sync
FORM ASC
Selects output format
ADEM:AM:RES? AVER
Reads RF time domain average results
ADEM:AM:RES? MAXH
Reads RF time domain max hold results
ADEM:AM:RES? MINH
Reads RF time domain min hold results
Retrieving Results