Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
182
User Manual 1173.9240.02 ─ 17
INPut:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
after an overload condition occured and the protection mechanism intervened.The
error status bit (bit 3 in the
STAT:QUES:POW
status register) and the
INPUT OVLD
message in the status bar are cleared.
(For details on the status register see the R&S
FSW User Manual).
The command works only if the overload condition has been eliminated first.
Usage:
Event
INPut:CONNector
<ConnType>
Determines whether the RF input data is taken from the RF input connector or the
optional Analog Baseband I connector. This command is only available if the Analog
Baseband interface (R&S
FSW-B71) is installed and active for input.
For more information on the Analog Baseband Interface (R&S
FSW-B71) see the
R&S
FSW I/Q Analyzer and I/Q Input User Manual.
Parameters:
<ConnType>
RF
RF input connector
AIQI
Analog Baseband I connector
*RST:
RF
Example:
INP:CONN:AIQI
Selects the analog baseband input.
Usage:
SCPI confirmed
Manual operation:
See
INPut:COUPling
<CouplingType>
This command selects the coupling type of the RF input.
Parameters:
<CouplingType>
AC
AC coupling
DC
DC coupling
*RST:
AC
Example:
INP:COUP DC
Usage:
SCPI confirmed
Manual operation:
See
Configuring the Measurement