Data management
R&S
®
FSVA3000/ R&S
®
FSV3000
644
User Manual 1178.8520.02 ─ 08
Remote command:
File Name
Contains the name of the data file without the path or extension.
By default, the name of a user file consists of a base name followed by an underscore.
Multiple files with the same base name are extended by three numbers, e.g.
limit_lines_005
.
File names must be compatible with the Windows conventions for file names. In partic-
ular, they must not contain special characters such as ":", "*", "?".
For details on the filename and location, see
Chapter 9.3.2.2, "Storage location and fil-
Comment
An optional description for the data file. A maximum of 60 characters can be displayed.
Remote command:
File Explorer
Opens the Microsoft Windows File Explorer.
Remote command:
not supported
File Type
Determines whether the global instrument settings with all "Channel"s are stored or
recalled, or the current "Channel" settings only.
Items:
Defines which data and settings are stored or are recalled. Depending on the "File
Type", either channels only, or global settings are available. Which items are available
also depends on the installed options (see also
Chapter 9.3.2.1, "Stored data types"
Remote command:
MMEMory:SELect[:ITEM]:HWSettings
MMEMory:SELect[:ITEM]:LINes:ALL
MMEMory:SELect[:ITEM]:TRACe<1...3>[:ACTive]
MMEMory:SELect[:ITEM]:TRANsducer:ALL
Save File
Saves the settings file with the defined filename.
Note:
Secure user mode. In secure user mode, settings that are stored on the instru-
ment are stored to volatile memory, which is restricted to 256
MB. Thus, a "memory
Storing and recalling instrument settings and measurement data