
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1241
User Manual 1178.8520.02 ─ 08
CALCulate<n>:LIMit<li>:UPPer:OFFSet
<Offset>
This command defines an offset for a complete upper limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Offset>
Numeric value.
*RST:
0
Default unit: dB
Manual operation:
See
CALCulate<n>:LIMit<li>:UPPer:SHIFt
<Distance>
This command moves a complete upper limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Distance>
Defines the distance that the limit line moves.
The unit depends on
Manual operation:
See
CALCulate<n>:LIMit<li>:UPPer:SPACing
<InterpolType>
This command selects linear or logarithmic interpolation for the calculation of an upper
limit line from one horizontal point to the next.
Suffix:
<n>
.
<li>
Parameters:
<InterpolType>
LINear | LOGarithmic
*RST:
LIN
Manual operation:
See
Analyzing measurements (basics)