Quick Start Guide 1313.9652.62 - 02
2.7
R&S FSQ
Preparing for Use
Preparing for Operation
2.1.6 Functional Test
➢
Call self alignment with the
CAL
key,
CAL TOTAL
softkey. Once the system correction
values have been calculated successfully, the message
Calibration Passed
will appear.
➢
Start the self test with the
SETUP
key,
SERVICE – SELFTEST
softkeys. Once the
instrument modules have been checked successfully, the message
Selftest Passed
will
appear.
Once both steps have been completed successfully, the instrument will be ready for operation.
The functional test should only be performed when the operating
temperature is reached (approx. 15 minutes after the instrument is
switched on).
The self alignment does not need to be repeated every time the
instrument is switched on, because the instrument saves the values
and loads them during booting automatically. A self alignment is
recommended if the instrument is used at a temperature that differs
considerable (more than 10°C) from the temperature present at the last
self alignment.
The self test also does not need to be repeated every time the
instrument is switched on. It is necessary only when instrument
malfunction is suspected.
Summary of Contents for FSQ series
Page 1: ...R S FSQ Signal Analyzer Quick Start Guide Quick Start Guide Test Measurement 1313 9652 62 02 ...
Page 28: ...R S FSQ 0 4 Quick Start Guide 1313 9652 62 02 ...
Page 50: ...R S FSQ Front and Rear Panel Rear Panel Connections 1 22 Quick Start Guide 1313 9652 62 02 ...
Page 110: ...R S FSQ Basic Operation Setting Parameters 4 12 Quick Start Guide 1313 9652 62 02 ...
Page 209: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix Appendix ...
Page 210: ...Quick Start Guide 1313 9652 62 02 R S FSQ Appendix ...
Page 220: ...R S FSQ Appendix Installing Non Plug Play Printers A 10 Quick Start Guide 1313 9652 62 02 ...