R&S FSH
Spectrum Analyzer Mode
Performing Spectrum Measurements
Operating Manual 1309.6275.12 - 14
64
Screen layout for the harmonic distortion measurement
1
Harmonics list
2
Total harmonic distortion in %
3
Total harmonic distortion in dB
4
Trace
5
Markers indicating harmonics (blue lines)
6
Harmonic distortion softkey menu
2.1.6.1 Defining the Number of Harmonics
By default, the R&S FSH shows the signal and its first harmonic. Each harmonic is
indicated by a marker that the R&S FSH places on the harmonic (here M1 and M2).
Note that all of the markers that have been set are normal markers that show the
absolute frequency of the harmonic.
At the same time, the R&S FSH also calculates the values for the total harmonic
distortion (THD) and shows the results in a box above the trace diagram. The values
are output in % as well as dB.
If you want to see more than one harmonic, the R&S FSH can show up to six
harmonics.
M
Press the MEAS key.
M
Press the "Harmonics" softkey.
M
Enter the number of harmonics you'd like to see.