R&S FSH
Receiver Mode
Selecting the Measurement Mode
Operating Manual 1309.6275.12 - 14
164
Screen layout of the bargraph result display
1
Measurement mode
2
Operating mode
3
Header table
4
Receive frequency and corresponding power level (numerical)
5
Bargraph
6
Receiver softkey menu
6.1.1.1 Defining the Receive Frequency
In single frequency measurement mode, the R&S FSH determines the power level of a
single frequency only. The available frequency range depends on the R&S FSH model
you are using.
M
Press the FREQ key.
M
Press the "Freq" softkey.
Enter the frequency you want to measure.
For a quick change of frequencies with the rotary knob, define a frequency stepsize.
M
Press the "Freq Stepsize" softkey.
Enter a frequency stepsize.
For measurements on systems that use channels instead of single frequencies, you
can also load a channel table and measure the channel power.
M
Press the "Channel Mode" softkey.
The R&S FSH opens a dialog box to select a channel table.
For measurements based upon a channel table, you can select a single channel
instead of a single receive frequency.
M
Press the "Channel" softkey.
M
Enter a channel number.