R&S FSH
Network Analyzer Mode
Performing Vector Measurements (R&S FSH-K42)
Operating Manual 1173.6275.12 - 30
195
6.4.2 Measuring the Reflection
The example included here is based on a reflection measurement of the two-port filter
that was also used for the transmission measurement.
You can skip the preset, vector measurement selection and frequency settings if you
have already set up the R&S FSH for the measurement. You can also skip the
calibration, if you have performed the appropriate calibration already.
Preset the R&S FSH
Before starting the measurement procedure, preset the R&S FSH to restore the default
configuration and connect the filter between the measurement ports.
► Press the PRESET key.
► Connect the DUT.
Start vector measurements.
► Press the MEAS key.
► Press the "Meas Mode" softkey.
► Select the "Vector" menu item.
Select the type of reflection measurement
► Press the "Result Display" softkey.
► Select the "Reflection Port 1" menu item for reflection measurements on port 1 or
the "Reflection Port 2" menu item for reflection measurements on port 2.
► Select the "Reflection on Port 1 and Port 2" menu item for simulatenous display of
the reflection on both ports.
For more information see "
Performing Scalar Measurements
Simultaneous display of several results
You can also display two results simultaneously in a single diagram. In that case, the
R&S FSH shows two traces with a different color in the result display.
See “
” for more information.