R&S FSH
Network Analyzer Mode
Calibrating Measurements
Operating Manual 1173.6275.12 - 30
184
● Normalize…
Normalization is a simple way to calibrate the measurement using one calibration
standard only. The correction data is deducted from this measurement. As only
one calibration standard is used, the isolation between the test ports is ignored. A
possible cross-talk between the test ports is therefore not eliminated and the
accuracy is lower than that of a full calibration.
Each calibration method is suitable for certain kinds of measurement tasks.
● For high isolation meaurements, use the normalization and isolation calibration
methods.
● For filter measurements requiring high standards (for example S11 in passband
better than -20 dB), use the full 2-port high accuracy calibration.
● For full s-parameter measurements on attenuators or amplifiers, use the full 2-port
calibration.
6.2.1.3 Performing Calibration
The procedure below shows a full 2-port calibration routine. All other calibration
methods basically work the same way, except for the type and number of calibration
standards you are going to need.
► Disconnect the DUT from the RF cable.
After disconnecting the DUT, the R&S FSH is ready for calibration.
► Press the MEAS key.
► Press the "Calibration" softkey.
► Select the "Full 2-Port" menu item.
The R&S FSH asks you to confirm
the currently selected calibration
kit.
If you are using another calibration kit, cancel the process and select the right one.
For more information see "
► Else, press the "Continue" softkey.
The R&S FSH asks you to connect
an "Open" first to port 1, then to
port 2.
► Firmly connect the "Open" of the calibration standard to the ports.
► You can abort the calibration any time by pressing the "Cancel" softkey.
► Press the "Continue" softkey to
start calibration.
The R&S FSH calibrates the open.