![Rohde & Schwarz FPL1000 Series User Manual Download Page 713](http://html.mh-extra.com/html/rohde-and-schwarz/fpl1000-series/fpl1000-series_user-manual_1477826713.webp)
Remote Commands
R&S
®
FPL1000
713
User Manual 1178.3370.02 ─ 03
If you change the horizontal position of marker <m2>, marker <m1> changes its hori-
zontal position to the same value.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:MARK4:LINK:TO:MARK2 ON
Links marker 4 to marker 2.
Manual operation:
See
CALCulate<n>:MARKer<m>[:STATe]
<State>
This command turns markers on and off. If the corresponding marker number is cur-
rently active as a delta marker, it is turned into a normal marker.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:MARK3 ON
Switches on marker 3.
Manual operation:
See
See
See
CALCulate<n>:MARKer<m>:TRACe
<Trace>
This command selects the trace the marker is positioned on.
Note that the corresponding trace must have a trace mode other than "Blank".
If necessary, the command activates the marker first.
Suffix:
<n>
.
<m>
Parameters:
<Trace>
1 to 6
Trace number the marker is assigned to.
Setting Basic Spectrum RF Measurement Parameters