Remote Commands
R&S
®
FPL1000
691
User Manual 1178.3370.02 ─ 03
In case of continuous sweep mode, the application calculates the moving average over
the average count.
In case of single sweep mode, the application stops the measurement and calculates
the average after the average count has been reached.
Suffix:
<n>
.
irrelevant
Parameters:
<AverageCount>
If you set an average count of 0 or 1, the application performs
one single sweep in single sweep mode.
In continuous sweep mode, if the average count is set to 0, a
moving average over 10 sweeps is performed.
Range:
0 to 200000
*RST:
0
Manual operation:
See
See
[SENSe:]AVERage<n>[:STATe<t>]
<State>
This command turns averaging for a particular trace in a particular window on and off.
Suffix:
<n>
.
<t>
Parameters:
<State>
ON | OFF | 1 | 0
[SENSe:]AVERage<n>:TYPE
<Mode>
This command selects the trace averaging mode.
Suffix:
<n>
.
Parameters:
<Mode>
VIDeo
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
*RST:
VIDeo
Example:
AVER:TYPE LIN
Switches to linear average calculation.
Setting Basic Spectrum RF Measurement Parameters