Remote Commands
R&S
®
FPL1000
680
User Manual 1178.3370.02 ─ 03
Example:
PMET2:TRIG:DTIMe 0.001
[SENSe:]PMETer<p>:TRIGger:HOLDoff
<Holdoff>
This command defines the trigger holdoff for external power triggers.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Holdoff>
Time period that has to pass between the trigger event and the
start of the measurement, in case another trigger event occurs.
Range:
0 s to 1 s
Increment: 100 ns
*RST:
0 s
Example:
PMET2:TRIG:HOLD 0.1
Sets the holdoff time of the trigger to 100 ms
Manual operation:
See
[SENSe:]PMETer<p>:TRIGger:HYSTeresis
<Hysteresis>
This command defines the trigger hysteresis for external power triggers.
The hysteresis in dB is the value the input signal must stay below the IF power trigger
level in order to allow a trigger to start the measurement.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Hysteresis>
Range:
3 dB to 50 dB
Increment: 1 dB
*RST:
0 dB
Example:
PMET2:TRIG:HYST 10
Sets the hysteresis of the trigger to 10 dB.
Manual operation:
See
[SENSe:]PMETer<p>:TRIGger:LEVel
<Level>
This command defines the trigger level for external power triggers.
Suffix:
<p>
.
1...4
Power sensor index
Setting Basic Spectrum RF Measurement Parameters