Remote Commands
R&S
®
FPL1000
676
User Manual 1178.3370.02 ─ 03
Parameters:
<Frequency>
The available value range is specified in the data sheet of the
power sensor in use.
*RST:
50 MHz
Example:
PMET2:FREQ 1GHZ
Sets the frequency of the power sensor to 1 GHz.
Manual operation:
See
[SENSe:]PMETer<p>:FREQuency:LINK
<Coupling>
This command selects the frequency coupling for power sensor measurements.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Coupling>
CENTer
Couples the frequency to the center frequency of the analyzer
MARKer1
Couples the frequency to the position of marker 1
OFF
Switches the frequency coupling off
*RST:
CENTer
Example:
PMET2:FREQ:LINK CENT
Couples the frequency to the center frequency of the analyzer
Manual operation:
See
[SENSe:]PMETer<p>:MTIMe
<Duration>
This command selects the duration of power sensor measurements.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Duration>
SHORt | NORMal | LONG
*RST:
NORMal
Example:
PMET2:MTIM SHOR
Sets a short measurement duration for measurements of station-
ary high power signals for the selected power sensor.
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters